    KLD Mono&Multiple Water Quality Analyzer
    On-line Analyzer for Monitoring Total Phosphorous in Water
  • On-line Analyzer for Monitoring Total Phosphorous in Water

        KLD-C-TP is an automatic on-line analyzer independently developed by KLD. Using imported CPU,highly precise signal processor and simple but steady loop flow system to effectively reduce signal fluctuation and to ensure the accuracy and stability of data. Complete separation between electronics and hydraulics allows long-term & stable running. It is suitable for several types of water matrix like river water, surface water and industrial effluent.


    Hits:157
    Product Presentation

        KLD-C-TP is an automatic on-line analyzer independently developed by KLD. Using imported CPU,highly precise signal processor and simple but steady loop flow system to effectively reduce signal fluctuation and to ensure the accuracy and stability of data. Complete separation between electronics and hydraulics allows long-term & stable running. It is suitable for several types of water matrix like river water, surface water and industrial effluent.

    KLD Mono&Multiple Water Quality Analyzer-5.jpg

    Features

    1. Robust & Reliable

        Designed for on-line industrial and environmental applications.Complete separation between electronics and hydraulics allows long-term stable operations.

    2. Precise Reagent Dosage

        CPU controlled photoelectric dosing system for accurate reagent dosage. Simple & steady structure ensures long-term service life.

    3. Easy Installation

        Delivered after a long & successful series of factory tests, it is ready for installation and setup. To start monitoring just need to connect reagent tubes, sample tube, water tube, waste tube and power supply, then set up the parameters.

    4. Automatic Calibration

        The analyzer performs calibration according to the calibration interval and type set up by users. If the result exceeds selected limits, the original calibration parameters will not be updated and an alarm signal will be outputted.

    5. Automatic Monitoring

        Measuring interval can be freely set up by users. Between two measurements the analyzer remains in standby.


    Benefits

  •  Automatic monitoring system gives alarm signal in abnormal condition. Easy and low maintenance;Low operating cost.

  •  4-20mA analog signal output; standard RS232, RS485 digital signal output, MODBUS protocol.

  •  New signal processor reduces signal fluctuation to ensure the accuracy and stability of data.

  •  Highly precise peristaltic pump extends the service life of pump tubes.

  •  Automatic restart when normal power supply after accidental power down.

  •  Remote control with local or remote PC connection.

  •  Low reagent consumption; short preparation time.

  •  Automatic diagnosis function. Identification and alarm for reagent shortage.

  •  Instrument operating signal output function.

  •  3000 groups of data storage and USB download for analysis data and running log.

  •  Easy operation, no special training is required.

  •  Controllable running of water pump and free setup for the sampling time of the pump.

  •  Security management function, automatic record and storage for all control operations. Read-only storage.

  •  2-level operation & management authorization: The system administrator can have all system operation and general operators can only perform routine maintenance & operation.


  • Measuring Principle

        The sample after proper filtration is pumped into the reactor, where injected with acid reagent and heated to 95°C, to convert all inorganic phosphides into orthophosphate. Through adding oxidation reagents and passing through a UV digestion step, to convert the rest of organic forms of phosphide into orthophosphate. Then, injecting antimony potassium tartrate, sodium molybdate and ascorbic acid into the sample. After fully mixing and reaction, the analyzer measures ABS value of blue substances produced by reaction at the specific wavelength, then calculates sample concentration according to the stored calibration data.

    Specifications

    Measuring Principle

    High Temperature Oxidation & UV Digestion, Colorimetric, Molybdate &Ascorbic Acid Method

    Operating Management
    Security management and 2-Level operation management function
    Measuring Type
    Manual or automatic with interval
    Measuring Interval
    Programmable; reverse control available by RS232
    Measuring Time
    45 minutes
    Measuring Range
    0.01~1mg/L, other ranges available on request
    Calibration Type
    Manual or automatic with interval
    Precision
    ≤ 5%
    Accuracy
    ±5%
    Zero Drift
    5%F.S
    Range Drift
    5%F.S
    Signal Output

    Standard 4-20mA analog output,Standard RS232/RS485 digital output, MODBUS Protocol

    Alarm

    High limit alarm, low limit alarm, reagent shortage alarm, calibration error alarm (standard)

    Sample and Waste Delivery
    Pressure free
    Reagent Temperature
    10-30°C
    Reagent Replacement
    3~4 weeks (depending on the ambient temperature)
    Ambient temperature
    5-40°C
    Protection
    IP55
    MTBF
    >720h
    Display

    Color touch screen (language: Chinese Simplified, Chinese Traditional, +English)

    Power supply
    220V AC±10%; power 90W
    Weight
    60kg (without reagents)
    Dimension
    1600×500×400mm(H×W×D) (with the height of base)

    Optional Functions

    Dual-channel

    One analyzer to monitor on-line for 2 outlets.

    Reagent Cooling

    Longer reagent service life and less reagent replacing.

    Alarm Function

    DI water low limit alarm and waste liquid high limit alarm function.

    Automatic Dilution Funccttiion

    Automatic dilution function for high concentration sample.

    Mini Printer

    Print data after measurement.